• • Integrated High Voltage on hold circuit.
• • External interlock disables HV output and defeats test when circuit is open.
• • Ergonomically designed for safety, speed, and efficiency.
• • Front panel indicates internal HV and polarity.
• • Polarity button selector on the front panel.
• • Microcontroller technologies.
• • Optional USB TestMinder, computer control.
• • Leads, manual and calibration certificate included.
• • One year calibration cycle
• • One year warranty.
Output Peak Voltage at 7KV: 6838 ±3%.
Open Circuit Rise Time: 9.13uS ±30% [Trise=1.67(T90%-T30%)] per IEC 1180.
Virtual Impedance: 40Ω.
Open Circuit Duration Time: 716uS ±20% (Time to half-value) per IEC 1180
Current Waveform: Not Defined.
Voltage Control: Manually, by turning a knob in the front panel or via USB.
Polarity Control: Positive and Negative. Alternating control available with USB.
Voltage Display: 4 Digit LED Display.
Voltage Meter resolution: 5V.
Voltage Meter Accuracy: ±1%.
Charge Time: <15 seconds.